Language Reference
CAL | Execute a calibration routine. |
Item | Description |
Preset State | CAL ALL |
Example
Description
10 OUTPUT 723;"CAL WL;"
The CAL command performs
The command arguments have the following denitions:
Argument | Denition |
CAL ALL | Calibrates for both wavelength and power. |
CAL PRESET | Erases all user calibration data. |
CAL PWR | Calibrates at the power level specied by the CALPWR |
| command at the entered calibration wavelength. The |
| maximum adjustment range is limited to 65 dB. The CAL |
| PWR command also veries that the calibration signal is |
| within 61 to 2 nm of the entered calibration wavelength. |
CAL WL | Calibrates at the wavelength specied by the CALWL |
| command. The maximum adjustment range is limited to |
| about 1 to 2 nm. The CAL WL command also veries |