Language Reference
BML | Subtract display line value from trace B. (B 0 DL 7 B) |
Example | 110 | OUTPUT | 723;"BLANK TRA;" | Blank trace A. |
| 120 | OUTPUT | 723;"SNGLS;" | Activate |
| 130 | OUTPUT 723;"CENTERWL 750NM;SP 100NM;" | Set measurement range . | |
| 140 | OUTPUT | 723;"DL | Set display line and reference |
|
|
|
| level. |
| 150 | OUTPUT | 723;"CLRW TRB;TS;" | Sweep trace B. |
| 160 | PAUSE |
| Press CONTINUE on computer |
|
|
|
| keyboard. |
| 170 | OUTPUT 723;"BML;" | Subtract display line from trace | |
|
|
|
| B. |
Description
See Also
The BML command subtracts the display line from trace B, then stores the dierence in trace B and places trace B in view mode. Use SNGLS and TS to be sure trace B contains a complete sweep of measurement information.
TRSTAT, TRPST, VIEW TRDSP, CLRW