Language Reference

MKCWL

MKCWL

Set center wavelength equal to wavelength at marker.

Example

40

OUTPUT

723;"TS;"

Sweep measurement range.

 

50

OUTPUT

723;"MKN EP;"

Move the marker using the keypad, step

 

 

 

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

 

 

 

keys, or knob. Press ACCEPT VALUE .

60

OUTPUT

723;"MKCWL;"

Move marker to center wavelength.

70

OUTPUT

723;"TS;"

Sweep.

Description

See Also

The MKCWL command centers the active marker and the trace point it marks by setting the center wavelength equal to the marker wavelength. If the active marker is a relative type that has been Activated by the MKD, MKAL, or MKAR commands, an error condition occurs. If no marker is displayed, MKCWL positions a marker at the center of the trace. If more than one trace is displayed, MKCWL positions it on the active trace or on trace A, B, or C, in that order; or marks the trace speci￿ed by MKTRACE. Update the trace data after executing MKCWL.

MKTRACE, MKOFF

7-265