Language Reference
MKCWL
MKCWL | Set center wavelength equal to wavelength at marker. |
Example | 40 | OUTPUT | 723;"TS;" | Sweep measurement range. |
| 50 | OUTPUT | 723;"MKN EP;" | Move the marker using the keypad, step |
|
|
| FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF |
|
|
| keys, or knob. Press ACCEPT VALUE . |
60 | OUTPUT | 723;"MKCWL;" | Move marker to center wavelength. |
70 | OUTPUT | 723;"TS;" | Sweep. |
Description
See Also
The MKCWL command centers the active marker and the trace point it marks by setting the center wavelength equal to the marker wavelength. If the active marker is a relative type that has been Activated by the MKD, MKAL, or MKAR commands, an error condition occurs. If no marker is displayed, MKCWL positions a marker at the center of the trace. If more than one trace is displayed, MKCWL positions it on the active trace or on trace A, B, or C, in that order; or marks the trace specied by MKTRACE. Update the trace data after executing MKCWL.
MKTRACE, MKOFF