Language Reference
STARTUP
STARTUP | Controls the start up conguration operation of the optical spectrum analyzer. |
Item | Description |
Default (no argument) | ONCE |
C A U T I O N | The STARTUP MAN and STARTUP ONCE commands both result in the | |
loss of any internal user calibration data when the system restarts. This is | ||
| ||
| because the data is part of the conguration. In addition, STARTUP MAN and | |
| STARTUP ONCE both force the POWERON command into IP if it was in LAST. | |
|
Example
Description
100 OUTPUT 723;"STARTUP AUTO;"
The rst time the the optical spectrum analyzer is turned on, it establishes and checks the communications links with the display. Thereafter, these conguration steps are skipped at turn on, unless the MMS conguration has changed.
Argument | Denition |
AUTO | The default operation which performs an automatic MMS |
| conguration the rst time the instrument is turned on. |
MAN | Forces the system to recongure at each power up. |
ONCE | Forces the system to recongure. After the reconguration, the |
| value of reverts to AUTO. |
For both STARTUP MAN and STARTUP AUTO, no eect is seen until the next power cycle. When the STARTUP ONCE command is executed, the