Language Reference
NORM
Example
Description
Query Response
Controls trace normalization for
100 OUTPUT 723;"NORM ON;"
The NORM command turns on trace normalization. When trace normalization is turned on, its operation depends on whether a logarithmic or linear display is selected. With logarithmic displays, AMB ON is executed, and the reference level is set to 0 dB. With linear displays, the reference level oset and display line are set to place the current reference trace at the top of the screen and set it equal to a value of 1.000X. Next, the ADBTL trace math function is turned on. This normalizes the current trace to 1.000X.
Turning o trace normalization simply disables the trace math enabled when NORM ON was executed. Note that turning o normalization in linear mode does not reset ROFFSET or DL to their previous values.
The query response value depends on the state of the NORM and MEASURE commands as shown in the following table:
NORM Query Response
Query Response
MEASURE State NORM OFF NORM ON
SA | 0 | 0 |
SR | 0 | 1 |