| LaserSpeed |
|
|
| |
|
|
| Quality Factor Current Value |
Q |
|
| |
Q<CR> | Read Only Command | ||
| Response Format: <N> | ||
| <N> is an integer from 0 to 15 indicating the current Quality Factor. A QF of 0 indicates that | ||
| the gauge is not currently making measurements, and a QF of 15 indicates that the gauge is | ||
| making excellent measurements. | ||
| This command returns the same value as you would normally receive from the TE data, or in | ||
| a response to the V command. | ||
|
|
| Reset Internal Length Count |
R |
|
| |
R<CR> or | Command | ||
R0<CR> |
|
|
|
| This command causes the | ||
| Pulse location to 0. |
Part No. 93463 / Drawing No. | Page 136 of 221 | Revision A (Sep 2007) |