LaserSpeed 8000-3 Instruction Handbook

Interfacing with the LS8000-3

RS-232 Transmit and Receive (Pins 1-2)

The LS8000-3 has an RS-232 serial port which can be used to receive measurement data, or to configure settings. The LS8000-3 will connect to any DB9 RS-232 port with a straight-through cable. A null modem is not required. Connections between the gauge and Industry standard pin outs for RS232 serial are given in the following tables:

RS-232 Serial Pinouts – LS8000-3 to Computer DB9 Connector

Gauge DB25 Pin

Gauge Signal

Computer DB9

Computer Signal

Number

Description

Pin Number

Description

1

RS-232

Transmit

2

RS-232

Receive

 

 

 

 

 

 

2

RS-232

Receive

3

RS-232

Transmit

 

 

 

 

11, 19, or 21

Signal Ground

5

Signal Ground

 

 

 

 

 

 

RS-232 Serial Pinouts – LS8000-3 to Computer DB25 Connector

Gauge DB25

Gauge Signal

Computer DB25

Computer Signal

Pin Number

Description

Pin Number

Description

1

RS-232

Transmit

3

RS-232

Receive

 

 

 

 

 

 

2

RS-232

Receive

2

RS-232

Transmit

 

 

 

 

11, 19, or 21

Signal Ground

7

Signal Ground

 

 

 

 

 

 

LaserTrak can be used to communicate with the LS8000-3 over the RS-232, RS-422, and Ethernet ports. Please see the RS232/RS422 Communication section for details on the Communication Protocol.

Part No. 93463 / Drawing No. 0921-01561