LaserSpeed 8000-3 Instruction Handbook
TE | Text Output Mode |
TE<CR> Data Output Command
Data Format: <L,V,QF,ST><CR>
This command places the
The
The measurement data is a comma delimited list of the current measurement data outputs. The response string is always 28 characters. All fields are
Format: |
|
|
|
|
|
±LLLLLLLLL,±VVVVVVVVV,QF,ST<CR> |
|
| |||
1234567890123456789012345678 |
|
| |||
| 1 | 2 |
|
|
|
|
|
|
|
|
|
Value | Format |
| Description | Examples |
|
L | Integer x 1000 |
| Current Length | +000001209 | (+1.209) |
| (Signed) |
|
| ||
V | Integer x 1000 |
| Current Velocity | +000120321 | (+120.321) |
| (Signed) |
|
| ||
QF | Integer |
| Quality Factor | 15 |
|
|
|
| 04 |
| |
|
|
|
|
| |
|
|
|
|
|
|
ST | Integer |
| Status | 63 |
|
|
|
| 47 |
| |
|
|
|
|
| |
|
|
|
|
|
|
The Status value is a bit field that has the following format:
Bit | Status Field | Convention |
0 | Laser @ Temperature | Yes=1 / No=0 |
|
|
|
1 | Laser Interlock Open/Closed | Closed=1 / Open=0 |
|
|
|
2 | Shutter Open/Closed | Open=1 / Closed = 0 |
|
|
|
3 | Material Present | Yes=1 / No=0 |
|
|
|
4 | Valid Measurement | Yes=1 / No=0 |
|
|
|
5 | System Ready (Indicates Interlock | Yes=1 / No=0 |
| Closed, Shutter Open, and Laser @ |
|
| Temperature) |
|
|
|
|
Part No. 93463 / Drawing No. | Page 155 of 221 | Revision A (Sep 2007) |