LaserSpeed 8000-3 Instruction Handbook
LaserTrak Software
Low Speed Pulse Config The low speed pulse output defaults to outputting pulse according to the Low Speed Pulse Rate set in the standard Configuration Tab. The pulse output can be configured to output gauge status if the low speed pulse output is not needed. The choices are shown in the table.
Length Threshold A | Indicates when the current length has exceeded the Length Threshold |
| A set point. This output is reset when the Index Pulse resets. See the |
| Index Pulse/Length Thresholds section for details about length |
| thresholds. |
|
|
Length Threshold B | Indicates when the current length has exceeded the Length Threshold |
| B set point. This output is reset when the Index Pulse resets. See the |
| Index Pulse/Length Thresholds section for details about length |
| thresholds. |
Advanced Operating Parameters
Direction Inversion Switch
Changes the function of the Direction Digital input.
When set to Normal, a low voltage on the Direction Input (or leaving it disconnected) will cause the gauge to measure a positive velocity and the length to count up, and a high
When set to Reverse, the polarity of the digital input is reversed, so a low will cause the gauge to measure a negative velocity, and a high
Changing the direction will also change the phase of the pulse outputs, causing a connected quadrature counter to count in the opposite direction.