LaserSpeed 8000-3 Instruction Handbook
# | Index Pulse Output Function |
#<CR> Read
Response Format: <N>
#<N><CR> Set
<N> is an integer specifying the function of the Index Pulse output. This command reconfigures the Index Pulse Output to be a status output.
Value Index Pulse Output Function
0(Default) Index (OPR) Pulse Output
1Material Present Output
2Valid Measurement Output
3QF Pulse Output
4QF Warning Output
5Shutter Open
6Laser On
7System Ready
8Thermostat Output (on @ 40C, off @ 35C)
9Light Stack Green
10 | Length Threshold A |
Material Present Output
This output reflects the Material Present status. It is the same as the Material Present bit in the status word of the serial output data.
Voltage High: Material is present
Voltage Low: Material is not present
Valid Measurement Output
This output indicates if the gauge is making valid measurements. The output is the same as the Valid LED on the gauge.
Voltage High: Valid Measurements are being made
Voltage Low: No measurements are being made
QF Pulse Output
This output is a frequency output that indicates the current Quality Factor. A frequency of 100Hz indicates a QF of 0, and a frequency of 1kHz indicates a QF of 15. The frequency of the output increases by 60Hz with each increase of 1 in the QF (100, 160, 220, 280, 340, …, 940, 1000)
Part No. 93463 / Drawing No. | Page 140 of 221 | Revision A (Sep 2007) |