LaserSpeed 8000-3 Instruction Handbook

Appendix E: Troubleshooting Guide

 

serial ports

you start LaserTrak

 

 

2. When LaserTrak starts, under the

 

 

Tools menu, click Program Settings

 

 

3. Place check marks next to only the

 

 

serial ports that should be searched

 

 

(usually only COM1.)

 

2. If using LaserTrak 3.x

1. Check data COMM LED to determine

 

the correct baud rate.

 

 

 

 

2. Start LaserTrak, but do not push the

 

 

search button. Push the OK button.

 

 

3. Go to Configuration, then to

 

 

Communication.

 

 

4. Adjust to the correct baud rate at the

 

 

real physical existing com port

 

 

(usually COM1) and press OK.

 

 

5. Push the read button to read the

 

 

configuration.

 

 

6. Start the visualisation and check that

 

 

the time on the right upper corner is

 

 

running correctly.

 

 

--- or ---

 

 

1. Start HyperTerminal and connect to

 

 

the COM port on which LaserTrak is

 

 

incorrectly detecting a gauge.

 

 

2. Start LaserTrak and have it search

 

 

for gauges. Since HyperTerminal

 

 

has the offending port already open,

 

 

LaserTrak won't be able to search it.

 

 

 

Settings changed on gauge

Switching between

When using LaserTrak, never read the

 

different gauges.

configuration of one gauge and update

 

 

the parameter settings to another gauge.

 

If communicating with

Enabling the settings lock will prevent

 

HyperTerminal, or with a

serial commands from changing settings.

 

PLC, settings may be

To lock the settings, send the command:

 

changed due to faulty

*LOCKEN=1<CR>

 

command.

 

When using a terminal program,

Gauge is initializing.

The LS9000/LS4000-3/LS8000-3 models

the gauge doesn't respond to

 

will respond to commands approximately

commands immediately after

 

5-10 seconds after power is applied. You

power-on.

 

can send a repeated Z<CR> command to

 

 

determine when the gauge is ready to

 

 

receive and respond to commands.

 

Incorrect Baud Rate

Make sure you are using the same baud

 

rate and framing as the gauge. At power

 

 

on, the gauge will auto-baud to whatever

 

 

commands you are sending, but it takes a

Part No. 93463 / Drawing No. 0921-01561

Page 207 of 221

Revision A (Sep 2007)