LaserSpeed 8000-3 Instruction Handbook

RS-232/RS-422 Communication

Indicates when the current length has exceeded the Length Threshold A set point. This output is reset when the Index Pulse resets. See the Index Pulse & Length Thresholds section for details about length thresholds.

Voltage High: Length Threshold A Exceeded

Voltage Low: Length Threshold A Not Exceeded

Y

Index (OPR) Pulse Length

Y<CR> Read

Response Format: <N>

Y<N><CR> Set

<N> is an integer from 0 to 255 that is the length of the Index (OPR) pulse in 10ms increments.

Example:

You want to set the OPR Pulse length to 180ms. Y = 180ms/10ms = 18.

Command: Y18<CR>

The time that the pulse is asserted does not vary with the measured velocity. When setting this value, be sure to set the pulse length to a value less than the time that will elapse between OPR pulses.

!

Index Pulse (OPR) Setting

!<CR> Read

Response Format: <N>

!<N><CR> Set

<N> is an integer specifying the number of User Scaleable Pulses per Index Pulse. Range: 1 to 999,999,999

Example:

Assume that the User Scaleable Pulse Rate (configurable with the P command) has been set to 1,000 pulses/foot.

To get an index pulse every 10 ft:

!= User Scaleable Pulse Rate (pulses/ft) x Desired Index Pulse Spacing (ft)

!= (1,000 pulses/ft) x 1 ft = 1,000 pulses

The command to set the correct value is !1000<CR>

To get an index pulse every 0.5 ft:

!= (1000 pulses/ft) x (0.5 ft) = 500 pulses Command to set the correct value: !500<CR>

Part No. 93463 / Drawing No. 0921-01561

Page 142 of 221

Revision A (Sep 2007)

Page 142
Image 142
Laser LS8000-3 manual Index OPR Pulse Length, Index Pulse OPR Setting, YCR Read, Yncr Set