Input and Output Calibration for Recorder with Display

5.4.2 Thermocouple Inputs Using a Compensated Calibrator

Refer to Figure 5-2and wire the recorder according to the procedure given in Table 5-5.

Table 5-5 Set Up Wiring Procedure for Thermocouple Inputs

Using a Compensated Calibrator

Step

Action

1Connect the thermocouple extension wire to the calibration source (see Figure 5-2).

2Connect the other end of the thermocouple extension wires to the TB2 terminals on the printed circuit assembly for the input channel being calibrated.

Main PCA

Main PCA

Pen 2

Pen 1

7

7

 

TB2

TB2

 

TB1

TB1

 

 

24825

 

 

TB2

 

 

+

 

 

 

 

R

 

 

-

Compensated

 

Calibrator

 

 

 

+

 

Thermocouple

 

 

extension wires

 

 

Figure 5-2 Calibration Set Up Diagram for Thermocouple Inputs

Using a Compensated Calibrator

12/03

DR4300 Circular Chart Recorder Product Manual