Step

Action/Result

Graphic

 

 

 

9

At the conclusion of the test turn off the power to the

 

 

recorder.

 

 

 

 

10

Being careful of the shock hazard at TB1, place run

 

 

switch SW5 to the run position (toward the bottom of the

 

 

board) on each PCA.

 

 

 

 

11

Set configuration switches on the PCA for each pen back

 

 

to the settings noted in Step 4.

 

 

 

 

12

Secure the chart plate in the operational position.

 

 

 

 

13

To resume normal operation leave SW5 in the run

 

 

position and power up the recorder.

 

 

 

 

Chart step pattern

Figure 3-4is a typical chart step pattern generated by the recorder in the step test.

24141

Figure 3-4 Typical Step Test Chart Patterns

68

DR4300 Circular Chart Recorder Product Manual

12/03

Page 80
Image 80
Honeywell manual Chart step pattern, DR4300 Circular Chart Recorder Step Action/Result Graphic