Installation

Main PCA

 

 

 

 

 

 

Main PCA

 

 

Pen 2

 

 

 

 

 

 

Pen 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

4

TB2

7

TB1

TB2

L

5 N

TB6

24821

2

3

 

 

1

It is the user’s responsibility to provide a switch and

 

 

 

non-time delay (North America), quick-acting, high

 

 

 

breaking capacity, Type F, (Europe) 1/2 A, 250 V

 

 

 

fuse(s) or circuit-breaker as part of the installation.

1

 

2

PROTECTIVE EARTH (Ground): Grounding of

 

 

 

this recorder shall be in accordance with National

 

 

 

and Local electrical codes.

L2/N

L1/H

3

NOISELESS (Clean) EARTH (Ground): To minimize

 

electrical noise and transients that may adversely

100 Vac to 240 Vac

 

 

affect the recorder, connection of the recorder terminal

(located in the rear of the case) to a local earth (ground), using a No. 11 (4 mm2) braided copper conductor, is required.

4For 2 Pen Recorder/Controller, connect ac power to TB1 of Pen 2 PCA. Supply connections must be made at TB6, if installed.

5 Wire ac power to this connector if installed.

Figure 2-9 Power Wiring – Models Without CE Mark

12/03

DR4300 Circular Chart Recorder Product Manual

29

Page 41
Image 41
Honeywell DR4300 manual Power Wiring Models Without CE Mark