A.2 Typical Reference Accuracy

Table A-1 Typical Reference Accuracy

Types of Input

 

Range

Reference

Factory

Temp Stability

Actuations

 

 

 

Accuracy*

Accuracy*

± Degrees

 

 

 

 

with

 

 

 

Error Per 1 Degree

 

 

 

 

Field

 

 

 

T

 

 

 

 

Calibration

 

 

 

 

°F

 

°C

± °F

 

± °C

± °F

± °C

 

 

 

 

 

 

 

 

 

 

 

Thermocouples

 

 

 

 

 

 

 

 

 

B

105 to 3300

 

41 to 1816

28.0

 

16.0

 

 

4.00

 

150 to 500

 

66 to 260

 

 

 

 

500 to 1000

 

260 to 538

6.0

 

3.3

37

21

1.00

 

1000 to 3300

 

538 to 1815

3.0

 

1.7

18

10

0.40

 

 

 

 

 

 

 

 

 

 

E

–454 to 1832

 

–270 to 1000

36.0

 

20.0

79

44

1.40

 

–454 to –202

 

–270 to –130

 

 

–202 to 1832

 

–130 to 1000

2.0

 

1.1

14

8

0.70

 

 

 

 

 

 

 

 

 

 

E (low)

–200 to 1100

 

–129 to 593

1.0

 

0.6

11

6

0.40

 

 

 

 

 

 

 

 

 

 

J

0 to 1600

 

–18 to 871

1.6

 

0.9

7

4

0.15

 

 

 

 

 

 

 

 

 

 

J (low)

20 to 770

 

–7 to 410

0.8

 

0.5

6

3

0.08

 

 

 

 

 

 

 

 

 

 

K

–320 to 2500

 

–196 to 1371

2.4

 

1.4

9

5

0.15

 

 

 

 

 

 

 

 

 

 

K (low)

–20 to 1000`

 

–29 to 538

1.0

 

0.6

7

4

0.10

 

 

 

 

 

 

 

 

 

 

Ni-Ni Moly

32 to 2500

 

0 to 1317

1.5

 

0.8

12

7

0.21

 

32 to 500

 

0 to 260

 

 

500 to 2500

 

260 to 1371

1.0

 

0.6

9

5

0.14

 

 

 

 

 

 

 

 

 

 

Nicrosil-Nisil

0 to 2372

 

–18 to 1200

2.1

 

1.2

11

6

0.20

 

 

 

 

 

 

 

 

 

 

R

0 to 3100

 

–18 to 1704

4.0

 

2.2

26

14

0.50

 

0 to 500

 

–18 to 260

 

 

500 to 3100

 

260 to 1704

2.0

 

1.1

13

7

0.21

 

 

 

 

 

 

 

 

 

 

S

0 to 3100

 

–18 to 1704

4.0

 

2.2

23

13

0.50

 

0 to 500

 

–18 to 260

 

 

500 to 3100

 

260 to 1704

2.0

 

1.1

13

7

0.21

 

 

 

 

 

 

 

 

 

 

T

–420 to 700

 

–251 to 371

1.6

 

0.9

13

7

0.15

 

 

 

 

 

 

 

 

 

 

T (low)

–200 to 600

 

–129 to 316

0.8

 

0.5

7

4

0.15

 

 

 

 

 

 

 

 

 

 

W5W26

0 to 4200

 

–18 to 2316

2.8

 

1.6

19

11

1.00

 

0 to 600

 

–18 to 316

 

 

600 to 3600

 

316 to 1982

2.6

 

1.5

13

7

0.50

 

3600 to 4200

 

1982 to 2316

3.2

 

1.8

27

15

0.20

RTD

 

 

 

 

 

 

 

 

 

Platinum

–300 to 900

 

–184 to 482

0.8

 

0.5

4

2

0.15

100 ohms**

 

 

100 ohms (low)**

–130 to 392

 

–90 to 200

0.6

 

0.3

3

2

0.15

100 ohms (T)***

–238 to 482

 

–150 to 250

0.6

 

0.3

3

2

0.15

 

 

 

 

 

 

 

 

 

 

*Includes reference junction calibration of ± 0.01 degrees using the standard "ice bath" method of calibration. Factory accuracy at reference ± 1.2 °F. Note that factory accuracy may have typical variations of ± 150 microvolts or ± 0.6 ohms for RTDs which means field calibration may be required to achieve stated accuracy.

**IEC Alpha = 0.00385

***Alpha = 0.00391