DR4300 Circular Chart Recorder

5.2 Input Calibration Minimum and Maximum Range Values

Minimum and maximum range values

You should calibrate the recorder for the minimum (0 %) and maximum (100 %) range values of your particular sensor. If your recorder has two (pens) inputs, calibrate each input separately.

Make sure the correct input is on display INP1 or INP2; use the

FUNC

key to select the input.

Select the voltage or resistance equivalent for 0 % and 100 % range values from Table 5-1.Use these values when calibrating your recorder.

142

DR4300 Circular Chart Recorder Product Manual

12/03