8.5.3 Pen Trace Troubleshooting

Procedure

Table 8-7explains how to troubleshoot a problem with the pen trace not matching the (correct) displayed value, assuming that the pen channel is configured to trace the input value.

Table 8-7 Troubleshooting Pen Trace Failure Symptoms

Step

What to do

How to do it or

 

 

where to find the instructions

 

 

 

1

Check the ink cartridge for proper

Reposition or replace the ink cartridge as

 

installation.

described in Section 6 – Routine Maintenance.

 

 

If the pen arm is severely warped, replace the

 

 

pen arm as described in Section 9 – Replacing

 

 

Hardware Assemblies.

 

 

 

2

Check that the chart agrees with the

Replace wrong chart with correct chart. Refer to

 

actuation type and check chart and pen

Section 4 – Configuration, Startup, and Operation

 

set up.

of Recorder with Display for instructions for

 

 

viewing the configuration.

 

 

See Section 6 – Routine Maintenance for

 

 

instructions for replacing the chart.

 

 

 

3

Check the pen alignment

Follow the procedure in Subsection 8.6.

 

 

 

4

Check the sensor for the proper type and

Verify the input configuration data and operation

 

its ability to function.

of the sensor.

 

 

Measure input signal at TB2 or apply calibration

 

 

source to verify input signal.

 

 

 

5

Replace the servo plate assembly.

Refer to the replacement procedure in Section 9

 

 

– Replacing Hardware Assemblies.

 

 

 

6

Replace the printed circuit assembly for

Refer to the replacement procedure in Section 9

 

the pen with the problem.

– Replacing Hardware Assemblies.

 

 

 

182

DR4300 Circular Chart Recorder Product Manual

12/03

Page 194
Image 194
Honeywell DR4300 manual Pen Trace Troubleshooting, Recorder with Display for instructions for