DR4300 Circular Chart Recorder

Lower Display

 

 

 

 

 

 

Indication

Type of Test

Reason For Failure

 

How to Correct the Problem

 

 

 

 

 

 

 

INP FL

background

Recorder detects an input failure.

1.

Be sure the range and actuation are

 

 

Two consecutive failures of input

 

configured correctly.

 

 

1 integration (for example,

2.

Check the input source.

 

 

recorder cannot make analog to

 

 

digital conversion.)

3.

See input set up group in Section 4 –

 

 

 

 

 

 

 

Configuration, Startup, and Operation of

 

 

 

 

Recorder with Display and change the “IN

 

 

 

 

TYP” or “LD CAL” to a different type.

 

 

 

 

Enter it, loop through the status tests,

 

 

 

 

then return the input type to the original

 

 

 

 

one.

 

 

 

4.

Recalibrate if step 3 does not clear the

 

 

 

 

error. Refer to Section 5 – Input and

 

 

 

 

Output Calibration for Recorder with

 

 

 

 

Display.

 

 

 

5.

Replace the printed circuit assembly for

 

 

 

 

the pen.

 

 

 

 

 

 

 

INPRNG

background

Input out of range. The process

1.

Make sure the range and actuation are

 

 

input is outside the range limits.

 

configured properly.

 

 

 

2.

Check the input source.

 

 

 

3.

See Section 4 – Configuration, Startup,

 

 

 

 

and Operation of Recorder with Display to

 

 

 

 

restore the factory or field calibration, as

 

 

 

 

applicable.

 

 

 

4.

Field calibrate using the procedure in

 

 

 

 

Section 5 – Input and Output Calibration

 

 

 

 

for Recorders with Display.

 

 

 

5.

Replace the printed circuit assembly for

 

 

 

 

the pen.

 

 

 

 

 

 

 

P1 ERR

power-up

Indicates that one or more error

 

 

 

 

1.

Press the

FUNC

key to toggle to the

P2 ERR

background

messages are active for the pen

 

other pen channel.

 

 

channel for which values are not

 

 

 

on display.

2.

Watch the display to see all error

 

 

 

 

 

For example, if value of INP1 is

 

messages for the active channel.

 

 

on display, while the input for pen

3.

Follow the procedure in this table for

 

 

2 fails, P2 ERR will be displayed.

 

 

 

each message displayed.

 

 

This alerts the operator to the

 

 

 

 

 

 

 

 

 

need to switch the display to pen

 

 

 

 

 

 

2 to view the active error

 

 

 

 

 

 

messages associated with that

 

 

 

 

 

 

channel’s input and printed circuit

 

 

 

 

 

 

assembly.

 

 

 

 

 

 

 

 

RAMTST

power-up

RAM failure

1. Cycle power or press the reset switch

 

 

 

 

SW2 to see if the error clears.

 

 

 

2.

If error does not clear, replace the printed

 

 

 

 

circuit assembly for the pen.

 

 

 

 

 

 

 

178

DR4300 Circular Chart Recorder Product Manual

12/03

Page 190
Image 190
Honeywell DR4300 manual P1 ERR, P2 ERR