Input and Output Calibration for Recorder with Display

Table 5-1 Voltage and Resistance Equivalents for 0 % and 100 % Range Values

Sensor Type

PV Input Range

 

 

Range Values*

 

 

 

 

 

 

 

 

 

Thermocouples

°F

 

°C

0 %

 

68 °F/20 °C

100 %

 

 

 

 

 

 

 

B

105 to 3300

 

41 to 1816

0 mV

 

13.769 mV

 

 

 

 

 

 

 

E

–454 to 1832

 

–270 to 1000

–9.835 mV

1.192 mV

76.373 mV

 

 

 

 

 

 

 

E (low)

–200 to 1100

 

–129 to 593

–6.472 mV

 

44.555 mV

 

 

 

 

 

 

 

J

0 to 1600

 

–18 to 871

–0.886 mV

1.019 mV

50.060 mV

 

 

 

 

 

 

 

J (low)

20 to 770

 

–7 to 410

–0.334 mV

 

22.400 mV

 

 

 

 

 

 

 

K

–320 to 2500

 

–196 to 1371

–5.822 mV

0.798 mV

54.856 mV

 

 

 

 

 

 

 

K (low)

–20 to 1000

 

–29 to 538

–1.114 mV

 

22.255 mV

 

 

 

 

 

 

 

Ni-Ni-Moly

32 to 2500

 

0 to 1371

–0.000 mV

0.755 mV

71.773 mV

 

 

 

 

 

 

 

Nicrosil-Nisil

0 to 2372

 

–17.8 to 1300

–0.461 mV

0.525 mV

47.513 mV

 

 

 

 

 

 

 

R

0 to 3100

 

–18 to 1704

–0.090 mV

0.111 mV

20.281 mV

 

 

 

 

 

 

 

S

0 to 3100

 

–18 to 1704

–0.092 mV

0.113 mV

17.998 mV

 

 

 

 

 

 

 

T

–420 to 700

 

–251 to 371

–6.187 mV

0.789 mV

19.097 mV

 

 

 

 

 

 

 

T (low)

–200 to 600

 

–129 to 316

–4.149 mV

 

15.771 mV

 

 

 

 

 

 

 

W5W26

0 to 4200

 

–18 to 2315

-0.234 mV

0.273 mV

37.075 mV

 

 

 

 

 

 

 

 

RTD

 

 

 

 

 

 

 

100 **

–300 to 900

 

–184 to 482

25.18

 

274.96

100 (low)**

–130 to 392

 

–90 to 200

64.30

 

175.84

100 (T)***

–238 to 482

 

–150 to 250

38.79

 

195.57

 

 

 

 

 

 

Milliamps

0 mA to 20 mA

0 mA

 

20 mA

 

 

 

 

 

 

4 mA to 20 mA

4 mA

 

20 mA

 

 

 

 

 

Millivolts

0 mV to 10 mV

0 mV

 

10 mV

 

 

 

 

 

 

0 mV to 100 mV

0 mV

 

100 mV

 

 

 

 

 

 

0 mV to 200 mV

0 mV

 

200 mV

 

 

 

 

 

 

 

Volts

 

0 V to 1 V

0 V

 

 

1 V

 

 

 

 

 

 

 

 

 

0 V to 2 V

0 V

 

 

2 V

 

 

 

 

 

 

 

 

 

1 V to 5 V

1 V

 

 

5 V

 

 

 

 

 

 

 

 

 

0 V to 5 V

0 V

 

 

5 V

 

 

 

 

 

 

 

0 V to 10 V

0 V

 

 

10 V

 

 

 

 

 

 

 

2 V to 10 V

2 V

 

 

10 V

 

 

 

 

 

 

 

 

*Range values for thermocouples are based on a cold junction temperature of 32 °F (0 °C). These values must be adjusted for the actual cold junction temperature when using the ambient temperature method for calibrating thermocouple inputs.

**IEC Alpha = 0.00385

***Alpha = 0.00391

12/03

DR4300 Circular Chart Recorder Product Manual

143

Page 155