Troubleshooting and Pen Alignment of Recorder without Display

7.3.4 Chart Rotation Troubleshooting

Procedure

Table 7-4explains how to troubleshoot a chart rotation problem.

Table 7-4 Troubleshooting Chart Rotation Failure Symptoms

Step

What to do

How to do it or

 

 

where to find the instructions

 

 

 

1

Check the chart installation.

Turn the chart until the desired time line on the

 

 

chart is aligned with the time index on the chart

 

 

plate and pen 1 before closing the locking clip on

 

 

the hub. Push down on the lifter to return the

 

 

pen(s) to the chart.

 

 

CAUTION: Do not try to rotate the hub

 

 

 

2

Check the settings of the configuration

Refer to Section 3 – Configuration, Startup, and

 

switches. Change settings if incorrect for

Operation of Recorder without Display.

 

application.

 

 

 

 

3

Run the step test.

Refer to Section 3 – Configuration, Startup, and

 

 

Operation of Recorder without Display

 

 

(Subsection 3.3.3).

 

 

 

4

Check the motor cable plug connection at

Power down the recorder, then visually examine

 

connector P3 on the printed circuit

the plug and reseat it.

 

assembly for pen 1.

 

 

 

 

5

Replace the chart motor.

Refer to the replacement procedure in Section 9

 

 

– Replacing Hardware Assemblies.

 

 

 

6

Replace the printed circuit board for pen

Refer to the replacement procedure in Section 9

 

1.

– Replacing Hardware Assemblies.

 

 

 