DR4300 Circular Chart Recorder

5.3 Input Calibration Preliminary Information

Equipment needed

Table 5-2lists the equipment required to calibrate various types of inputs. You will also need a medium size bladed screwdriver for general use and a small bladed screwdriver for use with the screws on the connector plugs.

Table 5-2 Equipment Needed for Calibration

Type of Input

 

Equipment Needed

 

 

 

Thermocouple Inputs

A compensated calibrator with ± 0.02 % accuracy.

(Compensated Calibrator)

Thermocouple extension wire that corresponds with the type of

 

 

 

thermocouple that will be used with the recorder input.

 

 

 

Thermocouple Inputs

A calibrating device with ± 0.02 % accuracy for use as a signal source

(Ice Bath)

 

such as a millivolt source.

 

Thermocouple extension wire that corresponds with the type of

 

 

thermocouple that will be used with the recorder input.

 

Two insulated copper leads for connecting the thermocouple

 

 

extension wire from the ice baths to the precision calibrator.

 

• Two containers of crushed ice, or an ice point reference.

 

 

 

RTD (Resistance Temperature

A resistance decade box, with ± 0.02 % accuracy, capable of

Detector)

 

providing stepped resistance values over a minimum range of 0 ohms

 

 

to 1400 ohms with a resolution of 0.01 ohm.

 

Three insulated copper leads for connecting the decade box to the

 

 

recorder.

 

 

 

Milliamp, Millivolt, and Volts

A calibrating device with ± 0.02 % accuracy for use as a signal source.

Inputs

• Two insulated copper leads for connecting the calibrator to the

 

 

 

recorder.

 

 

 

144

DR4300 Circular Chart Recorder Product Manual

12/03

Page 156
Image 156