DR4300 Circular Chart Recorder

Step

Description

Press

Action/Result

2

Calibrate 0 %

FUNC

You will see:

 

 

 

Upper Display

 

 

 

 

APLY

 

 

 

 

 

 

 

 

 

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

 

INZERO

 

 

 

 

 

 

 

 

 

 

 

• Adjust your calibration device to an output signal equal

 

 

 

 

to the 0 % range value for your particular input type. See

 

 

 

 

Table 5-1for Voltage or Resistance equivalents.

 

 

 

• Wait 60 seconds, then go to the next step.

 

 

 

 

3

Calibrate 100 %

 

You will see:

FUNC

 

 

 

Upper Display

 

 

 

 

APLY

 

 

 

 

 

 

 

 

 

 

 

Lower Display

 

 

 

 

INSPAN

 

 

 

 

 

Adjust your calibration device to an output signal equal to

 

 

 

the 100 % range value for your particular input type (see

 

 

 

Table 5-1for voltage or resistance equivalents). Wait 60

 

 

 

seconds, then go to next step.

 

 

 

 

4

Exit the Calibration Mode

 

The recorder stores the calibration constants.

FUNC

 

 

The value of input set up group parameter “LD CAL” is set

 

 

 

 

 

 

to “FLD” (field).

 

 

 

 

 

 

The recorder begins to use the new field calibration values

 

 

 

for this channel.

 

 

 

Repeat this procedure for the other pen, if required.

152

DR4300 Circular Chart Recorder Product Manual

12/03

Page 164
Image 164
Honeywell DR4300 manual Repeat this procedure for the other pen, if required