Routine Maintenance

6.3 Replacing the Ink Cartridge

Procedure

Refer to Figure 6-1and follow the procedure in Table 6-2to replace the ink cartridge.

Table 6-2 Procedure for Replacing the Ink Cartridge

Step

Action

1Remove the power from the recorder, then open the door.

2Pull up on the pen lifter to raise the pen(s) from the chart plate.

ATTENTION Be careful not to move the pen arm while removing and installing the ink cartridge.

3Unclip and remove the purple (pen 1) or red (pen 2) ink cartridge from the pen arm.

4Remove the protective cap from the pen tip on the new cartridge and open its clip. (If you plan to store or ship the recorder, save the protective cap.)

5Slide the new cartridge onto the pen arm so that its tip fits into the notch at the end of the pen arm and close the clip to secure the cartridge to the pen arm.

6Push down the pen lifter to return the pen tip to the chart.

7Close the door and apply power.

12/03

DR4300 Circular Chart Recorder Product Manual

159

Page 171
Image 171
Honeywell manual Procedure for Replacing the Ink Cartridge, DR4300 Circular Chart Recorder Product Manual 159