Troubleshooting and Pen Alignment

 

 

 

 

of Recorder with Display

 

 

 

 

 

 

 

Step

Press

Action/Result

 

 

 

 

 

 

8

 

To set tip of pen 1 (purple) on zero scale line of chart ± 0.25 % (or quarter

 

 

 

 

 

 

graduation).

 

 

 

 

 

 

 

 

or

 

 

 

 

 

 

 

 

 

 

 

 

9

FUNC

and you will see:

 

 

Upper Display

 

 

 

 

 

 

 

 

(value)

 

(approximately 1200 more than the zero value in Step 2)

 

 

 

 

 

 

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

 

 

 

PEN 100

 

 

 

 

 

 

 

 

 

Wait until pen stops moving upscale.

 

 

 

10

 

To set tip of pen 1 (purple) on full scale line of chart ± 0.25 % (or quarter

 

 

 

graduation).

 

 

 

 

 

 

 

 

or

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

DISP

to return to normal operation for pen 1.

 

you will see:

 

 

 

 

 

 

 

Upper Display

 

 

 

 

(value)

 

 

 

 

 

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

INP

and the Input 1 indicator will be lit.

 

 

 

 

 

 

 

 

 

 

If the recorder has 2 pens go to Step 12.

 

 

 

 

12

 

you will see:

 

FUNC

 

 

 

Upper Display

 

 

 

 

 

 

 

 

(value)

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

INP

and the Input 2 indicator will be lit.

 

 

 

 

 

 

 

 

 

 

Repeat Steps 1, 2, and 8 through 11 for pen 2 (red).