DR4300 Circular Chart Recorder

Millivolts, Volts and Milliamps inputs

Refer to Figure 5-5and wire the recorder according to the procedure given in Table 5-8.

Table 5-8 Set Up Wiring Procedure for Calibrating Millivolts, Volts, and Milliamps Inputs

Step

Action

1Connect the copper wire to the calibration source (see Figure 5-5).

2Connect the other end of the copper wire to the TB2 terminals on the printed circuit assembly for the input channel being calibrated.

Main PCA Pen 2

Main PCA Pen 1

7

7

TB2

TB2

TB1

TB1

TB2

+

R

24094

Calibration source

-

+

Copper wires

Figure 5-5 Calibration Set Up Diagram for Millivolts, Volts, and Milliamps Inputs

150

DR4300 Circular Chart Recorder Product Manual

12/03

Page 162
Image 162
Honeywell DR4300 manual Millivolts, Volts and Milliamps inputs