Configuration, Startup, and Operation of Recorder with Display

4.3.20 Lock Out Parameter Set Up Group

Introduction

This is the parameter used to lockout any unauthorized changes to the recorder’s configuration, calibration, and pen alignment parameters. Calibration group functions are described in Section 5 – Input and Output Calibration for Recorder with Display. Pen alignment group functions are described in Section 8 Troubleshooting and Pen Alignment for Recorder with Display.

Set this group last

Because this group contains functions that have to do with security and lockout, we recommend that you configure this group last, after all the other configuration data has been entered.

Lockout group prompts

Table 4-18lists the function prompt in the lock out set up group.

Press SETUP key until “LOCK” appears in the lower display.

Press FUNC key to select the parameter.

Table 4-18 Lockout Parameter Definitions

Lower Display

 

Upper Display

Parameter

Prompt

Range of Setting

Definition

 

 

or Selection

 

 

 

 

 

LOCK

 

 

LOCK limits access to one or more types of set up

 

 

 

groups. Do not configure until all configuration is

 

 

 

complete.

 

 

 

 

 

NONE

[factory setting]

NO LOCKOUT—All groups read/write.

 

 

 

 

 

CAL

 

CAL—Locks out calibration and pen alignment groups;

 

 

 

they will not even be displayed. All other groups are

 

 

 

available for read/write.

 

 

 

 

 

CONF

 

CONF—Tuning and cycle parameters are read/write. All

 

 

 

other configuration groups (except lock) are read only.

 

 

 

Calibration and pen alignment groups are not displayed.

 

 

 

 

 

FULL

 

FULL— All configuration groups (except lock) are read

 

 

 

only. Calibration and pen alignment groups are not

 

 

 

displayed.

 

 

 

 

ATTENTION

The Lock function cannot be changed when the RUN/SETUP switch is in the "SETUP" position, making it usable as a physical lockout. Whatever lockout level is selected, it is protected as long as the switch is in the SETUP position.

12/03

DR4300 Circular Chart Recorder Product Manual

115

Page 127
Image 127