DR4300 Circular Chart Recorder

Lower Display

Upper Display

 

Parameter

 

 

 

Prompt

Range of Setting

 

Definition

 

 

 

 

or Selection

 

 

 

 

 

 

 

 

 

 

IN TYP

 

INPUT ACTUATION TYPE—This selection specifies what actuation

 

 

you are going to use for the input. Be sure that the values

 

 

configured for the high and low chart range, alarm setpoint, etc. are

 

 

within the measuring range for the selected actuation type.

 

B

B thermocouple

105

to

3300°F

41

to

1815°C

 

EH

E thermocouple high

–454

to 1832°F

–270

to 1000°C

 

EL

E thermocouple low

–200

to

1100°F

–129

to

593°C

 

JH

J thermocouple high

0

to

1600°F

–18

to

871°C

 

JL

J thermocouple low

20

to

770°F

–7

to

410°C

 

KH

K thermocouple high

–320

to 2500°F

–196

to 1371°C

 

KL

K thermocouple low

–20

to

1000°F

–29

to

538°C

 

NNM

NiNiMo thermocouple

32

to 2500°F

0

to 1371°C

 

NIC

Nicrosil-Nisil

0

to

2372°F

–18.8

to

1300°C

 

 

thermocouple

 

R

R thermocouple

0

to 3100°F

–18

to 1704°C

 

S

S thermocouple

0

to

3100°F

–18

to

1704°C

 

TH

T thermocouple high

–420

to

700°F

–251

to

371°C

 

TL

T thermocouple low

–200

to

600°F

–129

to

316°C

 

W

W thermocouple

0

to 4200°F

–18 to 2316°C

 

100 H*[factory setting]

100 Ohm–RTD (high)

–300 to 900°F

–184 to 482°C

 

100 L*

100 Ohm-RTD (low)

–130 to 392°F

–90 to 200°C

 

100 T**

100 Ohm-RTD (special) –238

to

482°F

–150

to

250°C

 

4-20***

4 to 20 Milliamps

 

 

 

 

 

 

* IEC Alpha =

0-20***

0 to 20 Milliamps

 

 

 

 

 

 

10m***

0 to 10 Millivolts

 

 

 

 

 

 

0.00385

100m***

0 to 100 Millivolts

 

 

 

 

 

 

**IEC Alpha =

 

 

 

 

 

 

200m***

0 to 200 Millivolts

 

 

 

 

 

 

0.00391

 

 

 

 

 

 

 

0-1***

0 to 1 Volt

 

 

 

 

 

 

***These input

0-2***

0 to 2 Volts

 

 

 

 

 

 

selections are not

1-5***

1 to 5 Volts

 

 

 

 

 

 

available on Limit

0-5***

0 to 5 Volts

 

 

 

 

 

 

Control units.

 

 

 

 

 

 

0-10***

0 to 10 Volts

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2-10***

2 to 10 Volts

 

 

 

 

 

 

 

 

 

XMITTR

 

Transmitter Input Type—Use this parameter to specify whether

 

 

the input should be used as the PV without characterization (linear)

 

 

or the square root of the input should be used as the PV.

 

Lin [factory setting]

Linear—No characterization applied to input.

 

 

 

 

Srt

Square Root—PV is the square root of input. For example:

 

 

suppose the linear input range is 0 to 5000 and the input type is 0 to

 

 

5 volts. Then if input = 2.5 V = 50 %,

 

 

 

.50 = .7071= 70.71%

.7071 x 5000 = 3535 = PV