Troubleshooting and Pen Alignment of Recorder with Display

8.3.2 View Status of Tests

Procedure

The status of self-diagnostic tests can be viewed using the procedure in Table 8-3.

Note that “FAILSF” is not a test. When “FAILSF YES” is displayed, it is an indication that the recorder is using the value of the control set up group “FAILSF” parameter as the output value because the recorder failed a self-test, or because the recorder detected an input failure while the input set up parameter “BRNOUT” = “NONE”.

See Table 8-4for more information about self-diagnostics, and for recommended actions when a test fails.

Table 8-3 Procedure for Displaying the Results of Self-Diagnostics

Step

Operation

Press

 

 

 

Action/Result

 

 

 

 

 

 

 

1

Select STATUS

 

Until you see:

 

 

 

SET

 

 

 

 

set up Group

 

Upper Display

 

 

 

 

UP

 

 

 

 

 

 

 

 

READ

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STATUS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Read the status tests

 

Until you see:

 

 

 

 

 

 

 

 

results

FUNC

 

 

Upper Display

YES indicates

 

 

 

 

 

NO or YES

 

 

 

 

 

 

a failure

 

 

 

 

 

Lower Display

 

 

 

 

 

 

 

 

 

 

 

 

 

FAILSF

 

 

 

 

 

 

Successive presses of the

 

key will display

 

 

 

FUNC

 

 

 

 

 

the results of the status tests (PASS or FAIL) in the

 

 

 

 

 

following order:

RAMTST

CFGTST

CALTST

• Identify the problem and correct the failure as shown in Table 8-4.

12/03

DR4300 Circular Chart Recorder Product Manual

175

Page 187
Image 187
Honeywell DR4300 manual View Status of Tests, Procedure for Displaying the Results of Self-Diagnostics