Performance Tests
12. Test Port Input Impedance
Required
Equipment
For 50 Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
For 75 Analyzers
Cable, 75, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Calibration Kit, 75, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . Agil ent 85036B
Procedure
1Connect the equipment as shown in Figure 4-26.
Figure4-26. S11 1-Port Cal Test Setup
2Press PRESET, MENU, NUMBER of POINTS, 1601, x1.
3Only for analyzers without Option 006: Press START, 50, k/m.
4Press CAL, CAL KIT & STDS, SELECT CAL KIT and select the appropriate calibration
kit:
If your analyzer is 50, press CAL KIT: 7mm.
If your analyzer is 75, press CAL KIT: N 75W.
5Press RETURN, RETURN, CALIBRATE MENU, S11 1-PORT.
6Connect an open to reference test port 1, a s shown in Figure 4-26.
7Press FORWARD: OPEN.