4-45
Performance Tests
12. Test Port Input Impedance
Required 
Equipment 
For 50Ω Analyzers 
Cable, APC-7, 24-inch   . . . . . . . . . . . . . . . .  . . . . . . . . .  Agilent P/N 8120-4779
Calibration Kit, 7 mm  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
For 75Ω Analyzers 
Cable, 75Ω, Type-N, 24-inch  . . . . . . . . . . . .  . . . . . . . .  Agilent P/N 8120-2408
Calibration Kit, 75Ω, Type-N  . . . . . . . . . . . . . . . . . . . . . . . . . . Agil ent 85036B
Procedure
1Connect the equipment as shown in Figure 4-26. 
Figure4-26. S11 1-Port Cal Test Setup 
2Press PRESET, MENU, NUMBER of POINTS, 1601, x1. 
3Only for analyzers without Option 006: Press START, 50, k/m. 
4Press CAL, CAL KIT & STDS, SELECT CAL KIT and select the appropriate calibration 
kit: 
•If your analyzer is 50Ω, press CAL KIT: 7mm. 
•If your analyzer is 75Ω, press CAL KIT: N 75W. 
5Press RETURN, RETURN, CALIBRATE MENU, S11 1-PORT. 
6Connect an open to reference test port 1, a s shown in Figure 4-26. 
7Press FORWARD: OPEN.