5-14
Automated Verification Option 011
Step 2. Measurement Calibration
Figure5-8. System Verification Test Setup
2While the equipment is warming up, review the connector care information in
the Agilent 8702D Users Guide.
3Insert the verification kit disk into the analyzers disk drive.
4Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK.
5If you want a printout of the verification data for all the devices, press SYSTEM,
SERVICE MENU, TEST OPTIONS, RECORD ON.
If you switch the record function on, you cannot switch it off during the verifi-
cation procedure.
6Position the paper in the printer so that printing starts at the top of the page.
7Press SYSTEM, SERVICE MENU, TESTS, SYS VER TESTS, EXECUTE TEST.
8The analyzer displays, Sys Ver Init DONE, when the initialization
procedure is complete.
Do not press PRESET or recall another instrument state. You must use the in-
strument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1Press CAL, CAL KIT & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE