2-7
Automated Verification
Step 3. Device Verification
each S-parameter measurement. 
•If you switched the record function on, the Agilent 8702D measures all S-pa-
rameters (magnitude and phase) without pausing. Also, the Agilent8702D 
only displays and prints the pass/fail information for the  S-parameter mea-
surements that are valid for system verification. 
Figure2-4. Connections for the 20 dB Verification Device 
4When the Agilent8702D finishes all the measurements, connect the 50 dB 
attenuator (supplied in the verification kit), as shown in Figure 2-5. 
Figure2-5. Connections for the 50 dB Verification Device 
5Press ⇑, 29, x1, EXECUTE TEST, CONTINUE. 
6When all measurements are complete, replace the verification device with the 
verification mismatch, as shown in Figure 2-6. Be sure that you connect Port A