6-81
Performance Tests – Option 011
Performance Test Records - Option 011
Test 12. Receiver Magnitude Dynamic Accuracy (1 of 2)
Receiver
Input Power (dB)
Agilent
8496A
Setting
(dB)
Receiver
Channel A
(dB)
Agilent
8902A
(dB)
Magnitude
Value (dB) Spec. (dB) Meas.
Uncer. (dB)
Channel A
– 10 0 _________ _________ _________ ≤ 0.075 ± 0.010
– 20 10 _________ _________ _________ ≤ 0.059 ± 0.020
– 30 20 _________ _________ _________ ≤ 0.050 ± 0.020
– 40 30 _________ _________ _________ ≤ 0.050 ± 0.040
– 50 40 _________ _________ _________ ≤ 0.050 ± 0.060
– 60 50 _________ _________ _________ ≤ 0.058 ± 0.080
– 70 60 _________ _________ _________ ≤ 0.089 ± 0.140
– 80 70 _________ _________ _________ ≤ 0.240 ± 0.160
– 90 80 _________ _________ _________ ≤ 0.680 ± 0.180
– 100 90 ______ ___ _________ _________ ≤ 1.950 ± 0.200
– 110 100 _________ _________ _________ ≤ 5.200 ± 0.280
Channel B
– 10 0 _________ _________ _________ ≤ 0.075 ± 0.010
– 20 10 _________ _________ _________ ≤ 0.059 ± 0.020
– 30 20 _________ _________ _________ ≤ 0.050 ± 0.020
– 40 30 _________ _________ _________ ≤ 0.050 ± 0.040
– 50 40 _________ _________ _________ ≤ 0.050 ± 0.060
– 60 50 _________ _________ _________ ≤ 0.058 ± 0.080
– 70 60 _________ _________ _________ ≤ 0.089 ± 0.140
– 80 70 _________ _________ _________ ≤ 0.240 ± 0.160