6-31
Performance Tests Option 011
9. Receiver Input Crosstalk
A into B Crosstalk from 1 GHz to 3 GHz
38 Press START, 1, G/n, STOP, 3, G/n.
39 Press MEAS, INPUT PORTS, A/R, AVG, AVERAGING OFF, MENU, TRIGGER MENU, SINGLE.
40 At the end of the sweep, press DISPLAY, DATA MEMORY, DATA/MEM, MEAS, INPUT
PORTS, B/R.
41 Press AVG, AVERAGING FACTOR, 5, x1, AVERAGING ON, MENU, TRIGGER MENU, NUMBER
of GROUPS, 5, x1.
42 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
(Only for Option 006 Analyzers) A into B Crosstalk from 3 GHz to
4.5 GHz
43 Press START, 3, G/n, STOP, 4.5, G/n.
44 Press MEAS, INPUT PORTS, A/R, AVG, AVERAGING OFF, MENU, TRIGGER MENU, SINGLE.
45 At the end of the sweep, press DISPLAY, DATA MEMORY, DATA/MEM, MEAS, INPUT
PORTS, B/R.
46 Press AVG, AVERAGING FACTOR, 5, x1, AVERAGING ON, MENU, TRIGGER MENU, NUMBER
of GROUPS, 5, x1.
47 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
(Only for Option 006 Analyzers) A into B Crosstalk from 4.5 GHz to
6 GHz
48 Press START, 4.5, G/n, STOP, 6, G/n.
49 Press MEAS, INPUT PORTS, A/R, AVG, AVERAGING OFF, MENU, TRIGGER MENU, SINGLE.
50 At the end of the sweep, press DISPLAY, DATA MEMORY, DATA/MEM, MEAS, INPUT
PORTS, B/R.
51 Press AVG, AVERAGING FACTOR, 5, x1, AVERAGING ON, MENU, TRIGGER MENU, NUMBER
of GROUPS, 5, x1.
52 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.