3-6
Manual Verification
Step 3. Device Verification
Figure3-3.
Figure3-3. Connections for the 20 dB Verification Device
3Press CONTINUE to run the test.
4Press MENU, POWER, –2, x1, CAL, CORRECTION ON.
5Press COPY, PRINT MONOCHROME to obtain the graphical representation of the
measurement.
6Press COPY, LIST, PRINT ALL MONOCHROME if you prefer a tabular representation
of the measurement.
7Press SYSTEM, SERVICE MENU, TEST OPTIONS, CONTINUE TEST.
Ignore the error message, CAUTION: CALIBRATION REQUIRED.
8Press MENU, POWER, –2, x1, CAL, CORRECTION ON.
9Repeat Step 5 through Step 8 until you have made all the S-parameter
measurements for the verification device.
For each verification device, there are a total of eight measurements (magni-
For those S-parameter measurements with upper and lower point limits shown on the
Agilent8702D display, you will notice that when you press the CORRECTION ON softkey,
the measurement test status changes from FAIL to PASS.
If the test status remains FAIL after you press CORRECTION ON, refer to “In Case of
Difficulty” on page3-9.