4-57
Performance Tests
15. Test Port Receiver Magnitude Compression
7For every test port input power level, calculate the receiver phase dynamic
accuracy using the following formula:
Phase Dynamic Accuracy = SIN 1 (B)
8Write the value in the Results Value column of the Performance Test
Record.
Test Port 1 Phase Dynamic Accuracy
9Repeat Step 1 through Step 8.
15. Test Port Receiver Magnitude Compression
Perform this test to verify the compression/expansion magnitude levels of the
analyzers test port receiver samplers.
Required
Equipment
For 50 Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75 Analyzers
Cable, 75, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Procedure Test Port 2 Magnitude Compression
1Connect the equipment as shown in Figure 4-34.
Figure4-34. Test Port Magnitude Compression Test Setup