6-30
Performance Tests Option 011
9. Receiver Input Crosstalk
(Only for Option 006 Analyzers) B into A Crosstalk from 4.5 GHz to
6 GHz
30 Press START, 4.5, G/n, STOP, 6, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
31 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
A into B Crosstalk from 300 kHz to 1 GHz
32 Connect the equipment as shown in Figure 6-15.
Figure6-15. A into B Receiver Input Crosstalk Test Setup
33 Press START, 300, k/m, STOP, 1, G/n.
34 Press MEAS, INPUT PORTS, A/R, AVG, AVERAGING OFF, MENU, TRIGGER MENU, SINGLE.
35 At the end of the sweep, press DISPLAY, DATAMEMORY, DATA/MEM, MEAS, INPUT
PORTS, B/R.
36 Press AVG, AVERAGING FACTOR, 5, x1, AVERAGING ON, MENU, TRIGGER MENU, NUMBER
of GROUPS, 5, x1.
37 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.