4-26
Performance Tests
7. Test Port Input Frequency Response
Figure4-14. Setup for Power Meter Calibration on Test Port 2
33 Press START, 3, G/n, STOP, 6, G/n.
34 Press CAL, PWRMTR CAL.
35 Press LOSS/SENSR LISTS, CAL FACTOR SENSOR B. Repeat Step 10 to build a
calibration factor sensor table for the Agilent 8481A power se nsor.
36 Press DONE to exit the sensor calibration factor entries menu.
37 To select the Agilent 8481A power sensor, press USE SENSOR B.
38 Press RETURN, TAKE CAL SWEEP to start the power meter calibration.