6-43
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
18 Connect the equipment as shown in Figure 6-23.
Figure6-23. Channel B Magnitude Dynamic Accuracy Test Setup
19 On the Agilent 8702D Option 011:
aPress MENU, POWER.
bUsing the analyzer front panel knob, adjust the test port power for a reading
of 0.000 dB on the Agilent 8902A.
cWait for the reading to stabilize.
20 On the Agilent 8902A, press the blue shift key and SET REF.
21 On the Agilent 8702D Option 011: