4-43
Performance Tests
11. System Trace Noise (Only for Analyzers with Option 006)
3Press MARKER FCTN, MARKER MODE MENU, STATS ON to activate the instruments
statistic feature.
System Trace Noise for A/R Magnitude from 30 kHz to 3 GHz
4Press MEAS, Trans: E/E S12 REV.
5Press MENU, CW FREQ, 3, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
6When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
7Write the s.dev (standard deviation) value shown, which appears on the
analyzer display, on the Performance Test Record.
System Trace Noise for A/R Magnitude from 3 GHz to 6 GHz
8Press MENU, CW FREQ, 6, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
9When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
10 Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase from 3 GHz to 6 GHz
11 Press FORMAT, PHASE.
12 Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
13 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
14 Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase from 30 kHz to 3 GHz
15 Press MENU, CW FREQ, 3, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
16 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
17 Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for B/R Magnitude from 30 kHz to 3 GHz
18 Press MEAS, Trans: E/E S21 FWD, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
19 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
20 Write the s.dev (standard deviation) value, which appears on the analyzer