Manual Verification
Step 3. Device Verification
Figure3-5. Mismatch Device Verification Setup
17 Press TESTS, 30, x1, EXECUTE TEST, CONTINUE. 
Ignore the error message. 
18 Press MENU, POWER, –2, x1, CAL, CORRECTION ON. 
19 Repeat Step 5 through Step 8 until you have made all the S-parameter 
measurements for the verification device. 
20 When the Agilent8702D displays, Ver Dev 3 FAIL, proceed to the next step 
only if all S-parameter measurements show a PASS status under the same 
condition mentioned in Step 10. 
21 Connect the verification mismatch as shown in Figure 3-6. Notice that port B 
of the verification mismatch is now connected to the reference test port 1.