4-32
Performance Tests
8. Test Port Crosstalk
In Case of 
Difficulty 
WARNING These servicing instructions are for use by qualified personnel only. 
To avoid electrical shock, do not perform any servicing unless you  are 
qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose 
dangerous voltages. Disconnect the instrument from all voltage 
sources while it is being opened.
WARNING The power cord is connected to i nternal capacitors that may remain 
live for five seconds after disconnecting the plug from its power 
supply.
1Remove the instrument top cover. Using an 8 lb-inch torque wrench, verify that 
all semi-rigid cables connected to the sampler/mixer assemblies are tight. In 
addition tighten any loose screws on the sampler/mixer assemblies (A4/5/6) 
and the pulse generator assembly (A7). 
2Remove the instrument bottom cover. Refer to Figure4-19. Verify  that cables 
W1, W31 and W32 are tight. 
3Repeat this test.