6-25
Performance Tests Option 011
8. Receiver Phase Frequency Response
Figure6-12. Phase Frequency Response Test Setup
Phase Frequency Response from 300 kHz to 3 GHz
2Press PRESET, MENU, POWER, 10, x1.
3Only for Analyzers with Option 006: Press STOP, 3, G/n.
4Press FORMAT, PHASE, MARKER FCTN, MARKER MODE MENU, STATS ON.
5Press SCALE REF, .6, x1, ELECTRICAL DELAY. Turn the analyzer front panel knob to
vary the electrical delay until the trace is in the most linear horizontal position.
6Press MENU, TRIGGER MENU, SINGLE. Write the p-p value, which appears on the
analyzer display, on the Performance Test Record.
7Press CONTINUOUS, MEAS, INPUT PORTS, B/R, FORMAT, PHASE.
8Repeat Step 5 and Step 6.
9Press CONTINUOUS, MEAS, INPUT PORTS, A/B, FORMAT, PHASE.
10 Repeat Step 5 and Step 6.
Phase Frequency Response from 3 GHz to 6 GHz
11 Press START, 3, G/n, STOP, 6, G/n, CONTINUOUS.