
6-40
Performance Tests – Option 011
12. Receiver Magnitude Dynamic Accuracy
Figure6-21. Receiver Magnitude Dynamic Accuracy 
Required 
Equipment 
Measuring Receiver  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8902A
Step Attenuator, 110 dB  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8496A
Attenuator, 10 dB. . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Cable (4), 50Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4781
2-Way Power Splitter (2), 50Ω . . . . . . . . . . . . . . . . . . . . . . . . .Agi lent 11667A
Adapter (2), Type-N (m) to Type-N (m) . . . . . . . . . . . Agilent P/N 1250-1475
Procedure Channel A Magnitude Dynamic Accuracy 
1On the Agilent 8902A: 
aPress the blue shift key and INSTR PRESET, CLEAR. 
bPress the yellow shift key and TUNED RF LEVEL. 
cPress LOG/LIN to have the Agilent8902A displayed linear readings (dBm). 
dPress 32.1, SPCL. 
2Set the Agilent 8496A to 20 dB. 
3On the Agilent 8702D Option 011: 
aPress PRESET, MENU, CW FREQ, 30, M/µ. 
bPress NUMBER of POINTS, 51, x1.