6-40
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
Figure6-21. Receiver Magnitude Dynamic Accuracy
Required
Equipment
Measuring Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8902A
Step Attenuator, 110 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8496A
Attenuator, 10 dB. . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Cable (4), 50, Type-N, 24-inch . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4781
2-Way Power Splitter (2), 50 . . . . . . . . . . . . . . . . . . . . . . . . .Agi lent 11667A
Adapter (2), Type-N (m) to Type-N (m) . . . . . . . . . . . Agilent P/N 1250-1475
Procedure Channel A Magnitude Dynamic Accuracy
1On the Agilent 8902A:
aPress the blue shift key and INSTR PRESET, CLEAR.
bPress the yellow shift key and TUNED RF LEVEL.
cPress LOG/LIN to have the Agilent8902A displayed linear readings (dBm).
dPress 32.1, SPCL.
2Set the Agilent 8496A to 20 dB.
3On the Agilent 8702D Option 011:
aPress PRESET, MENU, CW FREQ, 30, M/µ.
bPress NUMBER of POINTS, 51, x1.