
6-29
Performance Tests – Option 011
9. Receiver Input Crosstalk
B into A Crosstalk from 300 kHz to 1 GHz 
22 Connect the equipment as shown in Figure 6-14.
Figure6-14. B into A Receiver Input Crosstalk Test Setup
23 Press START, 300, k/m, STOP, 1, G/n.
24 Press MEAS, INPUT PORTS, A/B, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 
25 When the analyzer finishes the number of sweeps, write the marker value, 
which appears on the analyzer display, on the “Performance Test Record.”
B into A Crosstalk from 1 GHz to 3 GHz 
26 Press START, 1, G/n, STOP, 3, G/n, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 
27 When the analyzer finishes the number of sweeps, write the marker value, 
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) B into A Crosstalk from 3 GHz to 
4.5 GHz
28 Press START, 3, G/n, STOP, 4.5, G/n, TRIGGER MENU, NUMBER of GROUPS, 5, x1. 
29 When the analyzer finishes the number of sweeps, write the marker value, 
which appears on the analyzer display, on the “Performance Test Record.”