4-54
Performance Tests
13. Test Port Receiver Magnitude Dynamic Accuracy
19 On the Agilent 8902A, press the blue shift key and SET REF.
20 On the Agilent 8702D:
aPress CAL, CALIBRATE MENU, RESPONSE, THRU.
bAfter the analyzer finishes the measurement, press DONE: RESPONSE.
cPress SAVE/RECALL, SAVE STATE.
21 Set the Agilent 8496A to 0 dB.
22 On the Agilent 8702D, press MENU, TRIGGER MENU, SINGLE.
23 Write the mean value in the Test Port column of the Performance Test
Record.
24 Write the Agilent 8902A readout in the 8902A column of the Performance
Test Record.
25 Calculate the difference between the Test Port 1" and the 8902A.
26 Take the absolute value of the calculated difference and enter the result in the
Calculated Value column of the Performance Test Record.
27 Set the Agilent 8496A to the other attenuation settings listed in the
Performance Test Record, while repeating Step 22 through Step 26.
NOTE
When the Agilent8902A displays the RECAL annunciation, press CALIBRATE.