4-64
Performance Tests
17. Test Port Output/Input Harmonics (Option002 Analyzers without Option 006 Only)
15 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 
16 After one sweep, press SCALE REF, AUTO SCALE. 
17 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 
18 Write the marker 1 value on the “Performance Test Record.” 
Port 1 Input Worst Case 2nd Harmonic 
19 Connect the equipment as shown in Figure 4-37. 
Figure4-37. Receiver Harmonics Test Setup 
20 Press PRESET, MENU, POWER, 8, x1. 
21 Press AVG, IF BW, 1, 0, x1. 
22 Press START, 16, M/µ, STOP, 1.5, G/n to set the frequency range. 
23 Press MEAS, Trans: E/E S12 REV, INPUT PORTS, A. 
24 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the 
trace. 
25 Press SYSTEM, HARMONIC MEAS, HARMONIC SECOND. 
26 After one sweep, press SCALE REF, AUTO SCALE to get a better viewing of the 
trace. 
27 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 
28 Write the marker 1 value (which appears on the analyzer display) on the 
“Performance Test Record.” This is the worst case port 1 input (receiver