6-4
Performance Tests Option 011
1. Source Frequency Range and Accuracy
Figure6-1. Source Frequency Range and Accuracy Test Setup
2Press PRESET, MENU, CW FREQ.
3Press 300, k/m and write the frequency counter reading on the "Performance
Test Record."
4Repeat Step 3 for each instrument frequency listed in the Performance Test
Record.
In Case of
Difficulty
1If any measured frequency is close to the specification limits, check the time
base accuracy of the counter used.
2If the analyzer fails by a significant margin at all frequencies (especially if the
deviation increases with frequency), the master time base probably needs
adjustment. In this case, refer to Frequency Accuracy Adjustment in the
Adjustments and Correction Constants chapter in the Agilent 875 3D Option
011 Service Guide. The Fractional-N Frequency Range Adjustment could
also affect the frequency accuracy.
3Refer to the Agilent8753 D Option 011 Service Guide for further
troubleshooting information.