4-30
Performance Tests
8. Test Port Crosstalk
Required
Equipment
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B
Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4779
For 75 Analyzers
Calibration Kit, 75, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . .Agil ent 85036B
Procedure
1Connect the equipment as shown in Figure 4-18.
Figure4-18. Test Port Crosstalk Test Setup
2Press PRESET, MENU, POWER, 10, x1.
3Press AVG, IF BW, 10, x1.