4-47
Performance Tests
12. Test Port Input Impedance
Figure4-28. S22 1-Port Cal Test Setup 
18 Press CAL, CALIBRATE MENU, S22 1-PORT. 
19 At the prompt, connect an open to reference te st port 2, as shown in 
Figure 4-28. 
20 Press REVERSE: OPEN. 
21 When the analyzer displays the prompt,  CONNECT STD THEN PRESS KEY 
TO MEASURE, connect a short to reference test port 2. 
22 Press REVERSE: SHORT. 
23 At the prompt, connect a load to reference test port 2. 
24 Press REVERSE: LOAD. 
25 When the analyzer displays, ’DONE’ IF FINISHED WITH CAL, press DONE 
1-PORT CAL. 
26 Press SAVE/RECALL, SAVE STATE to save the 1-Port calibration. 
27 Connect the equipment as shown in Figure 4-29.